Catheter Management of Left Atrial Appendage Closure Devices

Earn CME/CE in your profession:


Continuing Education Activity

For patients with atrial fibrillation, percutaneous left atrial appendage closure (LAAC) emerged as a valid alternative to oral anticoagulation. This activity reviews the evaluation, indications/contraindications, pre/post-procedural management of percutaneous LAAC and highlights the role of the interprofessional team in evaluating and treating this condition.

Objectives:

  • Identify indications and contraindications for percutaneous left atrial appendage closure.
  • Describe pre-procedural planning and preparation to perform successful percutaneous left atrial appendage closure.
  • Summarize the complications of percutaneous left atrial appendage closure.
  • Explain the importance of collaboration and communication amongst the interprofessional team to ensure the appropriate selection of candidates for percutaneous left atrial appendage closure and to enhance post-procedural management.

Introduction

Atrial fibrillation (AF) is the most common type of heart arrhythmia affecting 33.5 million people worldwide.[1][2] The most dreaded complication of this disease is stroke, which is also the leading cause of disability in the United States.[3] AF increases the risk of ischemic stroke by 4 to 5 fold in all ages when left untreated.[3][4] Moreover, AF is correlated with a higher risk of extracranial thromboembolic aortic events and also in the mesenteric, renal, and peripheral arteries.[5] The number of strokes attributed solely to AF increases with age approaching 23.5%.[4][6] In AF patients, oral anticoagulants (OACs) remain the gold standard treatment. The role of OACs in preventing strokes is well established, yet it is contraindicated in patients with an increased risk of bleeding.[7][8] Hence, left atrial appendage occlusion (LAAO) has risen as an alternative approach in this subset of patients.

Anatomy and Physiology

The left atrial appendage (LAA) is a trabeculated long tubular structure connected to the venous portion of the left atrium (LA) through a narrow junction. The size and shape vary widely, but in about 70% of the patients, the appendage has a bent or spiral axis.[9] LAA divides anatomically into three portions; the ostium, the neck, and the lobar region.[10] Remodeling of LAA leading to dilation, stretching, and reduction of pectinate muscle volume occurs in chronic AF patients.[11] On transesophageal echocardiography (TEE), about 5% to 15% of AF patients have atrial thrombi, which are situated in LAA in about 91% of non-rheumatic AF patients.[12][13][14][15][16] 

During the sinus rhythm, the contraction of LAA causes vigorous blood flow in and out of the appendage cavity. When there is a contractile dysfunction in the LAA, as in AF, it can predispose to local stasis, thrombosis, and systemic embolization. Changes in the left ventricular function (which may be impaired in AF) influence the extent of LAA filling and emptying more than the LAA function.[17] In normal conditions, the ventricular filling produces a suction effect inside the cavity that affects LA and LAA emptying and loading.

Indications

The majority of the institutions consider LAAC when AF patients are at high thromboembolic risk (with a CHADS2DS2-Vasc score of greater than or equal to 3 or CHADS2 score of greater than or equal to 2) with contraindication for long term oral anticoagulants (OACs), as well as patients with thromboembolic events despite adequate use of OACs. It is also indicated when there is a high probability of therapeutic non-compliance to OACs and patients with OAC intolerance. The 2019 American College of Cardiology (ACC) and the American Heart Association (AHA) guidelines on the management of atrial fibrillation gave a class IIb recommendation for percutaneous LAAC in patients with an increased risk of stroke who cannot tolerate long term anticoagulation.[18] There was a similar recommendation by the European Society of Cardiology (ESC) and the European Stroke Organization (ESO) in their 2016 guidelines for managing atrial fibrillation.[19] 

The depth of the left atrial appendage must be at least longer than the width of the ostium of the LAA for optimal device deployment and closure of LAA. The United States Food and Drug Administration (FDA) approved only one device for stroke and systemic embolism reduction in 2015, the Watchman (Boston Scientific, Natick, MA, USA). However, its use is limited to patients at increased stroke risk and can tolerate short-term anticoagulation. Another device used off-label for LAAC is the Lariat (SentreHeart Inc., Redwood City, CA, USA), which is FDA-approved for facilitating percutaneous ligation of LAA. Percutaneous LAAC is performed off-label for patients intolerant to OACs and at high risk of stroke. However, there is a lack of strong evidence supporting its use in these patients. In AF patients undergoing cardiac surgery, concomitant surgical ligation of the left atrial appendage is a class IIb indication.[20]

Contraindications

Patients with a low risk for stroke with CHADS2DS2-Vasc score less than 3 or a CHADS2 score of less than 2 are not recommended to have LAAC. Valvular heart disease, such as mitral stenosis, is another contraindication for the intervention. Patients with left atrial thrombus or tumor, active infection, the presence of atrial septal defect (ASD), or a prior patent foramen ovale (PFO) closure device are not recommended for trans-septal catheterization. Patients who have other indications for long-term or lifelong OACs, such as patients with a prosthetic valve, pulmonary embolism, deep vein thrombosis, and thrombi in the left atrium or ventricle, should not undergo left atrial appendage closure.[21] The left atrial appendage closure is also contraindicated if the depth of the LAA is significantly shorter than the width of the ostium of LAA. Based on current guidelines, patients who are unable to tolerate short-term oral anticoagulation are contraindicated to undergo percutaneous LAAC.

Equipment

PLAATO

In 2001, Sievert achieved the first-in-human percutaneous LAAC using the PLAATO device (Appriva Medical, Sunnyvale, CA).[22] The device had a self-expanding metal cage made of nitinol, coated with non-thrombogenic polytetrafluoroethylene, with anchors along three struts to help anchor the device into the LAA, and was available in different sizes from 15 mm to 32 mm. It was delivered through a 14-Fr delivery sheath guided by fluoroscopy and transesophageal echocardiography (TEE). The device was withdrawn from the market in 2006 despite a favorable clinical trial.[23]

Watchman and Watchman FLX

Watchman device (Boston Scientific, Natick, MA) is the only device that has been evaluated in randomized clinical trials.[24][25][26] It is composed of a 10-strut semi-spherical self-expanding nitinol frame with a 160 μm permeable polyethylene terephthalate membrane fabric cap facing into the body of LA. The device size is based on the diameter of the proximal shoulder of the device, which is its widest portion. The model comes in 21 mm, 24 mm, 27 mm, 30 mm, and 33 mm sizes. The open distal end is fixed in 1 row by ten active fixing anchors. The device is connected by a threaded insert inside the proximal cap to a delivery cable. The outer diameter of the transseptal access sheath is 14F and comes in anterior, double, and single curves. The double curve is used in more than 90% of the procedures. The device needs to be 8% to 20% larger than the body of LAA. This device is used to occlude LAAs about 17 mm to 30 mm in diameters as long as there is adequate depth to accommodate the device, the length of which is approximately equal to its fully expanded diameter. Before release from the delivery cable, the device can be partially recaptured and redeployed if the location of the implant is too deep, and it can also be fully recaptured and removed if a different device size is required or it is deployed too proximally. The device is authorized for use in the United States and is CE marked in the European Economic Area for sale.

The WATCHMAN FLX (Boston Scientific) is a second-generation device fully covered by the polyethylene terephthalate membrane cap to minimize peri-device leaks.[27] It is available in 5 sizes (20 mm, 24 mm, 27 mm, 31 mm, and 35 mm) for 15 mm to 32 mm LAA ostia and is 10% to 20% shorter than the previous generation unit. Combined with greater radial strength, the increased number of struts (18), and anchors (12 in 2 rows), it provides improved tissue fixation. While a greater compression (10% to 37%) was initially permitted, an increased risk of device embolization with excessive oversizing was observed. The WATCHMAN FLX got the CE-mark approval in 2015 but was removed from the European market in March 2016 due to increased implant embolization incidents.

AMPLATZER Cardiac Plug and Amulet

The AMPLATZER Cardiac Plug (ACP; Abbot Vascular, Santa Clara, CA) is a self-expanding nitinol mesh with a proximal polyester fabric disk and distal lobe connected by an articulated waist.[28] The device length is less than its diameter, accommodating to shorter LAAs. There are six pairs of stabilizing wires in the lobe aimed to be implanted in the orifice of LAA at 10 mm, while the disk is designed to seal the ostium at the left atrial side. Its delivery sheath (TorqVue TM) has two 45° curves to help with positioning. Eight different sizes based on the lobe dimensions, from 16 mm to 30 mm, are available (12.6 mm to 28.5 mm for landing zone measurements). If required, the ACP can be recaptured and re-positioned.

The Amulet device (Abbot Vascular) is based on ACP device design but with improvements facilitating implantation and reducing procedural complications.[28] Eight different sizes of this device ranging from 16 mm to 34 mm, are available, which fit LAA sizes of 11 mm to 31 mm (landing zone measurements). The distal lobe of the Amulet device is 2 mm to 3 mm longer, and the diameter of the proximal disk is 6 mm to 7 mm larger than the distal lobe diameter when compared to the ACP device and has more stabilizing wires and a longer waist improving stability. Furthermore, to reduce device thrombosis, the proximal end screw was recessed. As a result, less oversizing is needed for this device due to its more stable design.[29] Both the ACP and the Amulet have been approved with the CE mark.

WaveCrest

The WaveCrest device (Biosense Webster, Diamond Bar, CA) is a single-lobe nitinol device with polyurethane foam facing the LAA and an expanded polytetrafluoroethylene cover on the left atrial side and has 20 anchoring points.[30] The occluder and anchoring systems can be independently manipulated in the device, and the contrast can be injected either proximally through the delivery sheath or distally through the occluder, ensuring good sealing. In addition, the device does not require delivery sheath placement in LAA, as the device is intended for proximal positioning in the neck of LAA, making it particularly beneficial for small LAA anatomies. In 2013, the WaveCrest (Biosense Webster Inc.) earned CE mark approval.

Occlutech LAA Occluder

The Occlutech LAA Occluder (Occlutech International AB, Helsingborg, Sweden) is a self-expanding conical-shaped nitinol wire mesh and is anchored at the distal margin through the closed loops.[31] The polyurethane coverage of the device promotes sealing and endothelization. It received the CE mark in 2016.

LAmbre LAA Closure System

The LAmbre LAA Closure System (Lifetech Scientific Co, Ltd, Shenzhen, China) is another nitinol device with a left atrial cover and a distal self-expanding umbrella, protected by a double polyethylene terephthalate membrane and attached with a central articulating waist.[32] A double stabilization mechanism with eight distal hooks and eight proximal U-shaped anchors secures the device. With hooks recessed until deployment, the distal umbrella is the first to be deployed, and the proximal cover is then released. Two types of LAmbre devices, including the standard and the special address single- and double-lobe anatomy, respectively. The special type is especially useful for multilobed or small LAAs. It has a low profile (8F to 10 F) and is fully recapturable and repositionable. This device earned its CE-mark approval in 2016. 

Ultraseal LAA Closure Device 

Another self-expanding device is an Ultraseal LAA closure device (Cardia Inc, Eagan, MN), which is comprised of two sections: a soft distal bulb for anchoring the device and a 3-leaflet proximal sail for occlusion of LAA. Both are connected with a dual articulating joint, enabling multidirectional mobility and optimum adjustment of various ostium angles and shapes.[33] The device has 12 stabilizing hooks to prevent dislodgement. It is available in 9 bulb sizes ranging from 16 mm to 32 mm, and oversizing of 25% to 33% is usually recommended. The proximal sail is 6 mm larger than the distal bulb, with a covering of polyvinyl acetate foam, with diameters of 22 mm to 38 mm. Two available delivery sheaths range from 10F to 12F, namely a single curve (45 degrees) and a double curve (45 degrees –45 degrees).

Pfm Device

The Pfm device (Pfm Medical, Köln, Germany) is a nitinol frame device consisting of three parts: a primary distal anchor, a middle connector, and a proximal disk with a secondary anchor.[34] The middle connector has an adjustable length accommodating variable LAA lengths. The device consists of a 10F to 12F delivery sheath and is secured by a barbless anchor, minimizing perforation risk.

Epicardial Devices

Lariat

The Lariat (SentreHeart, Redwood City, CA) device enables the percutaneous delivery of pretied surgical sutures to litigate the LAA. It is cleared by the FDA for the approximation of soft tissue.[35] It is used off-label for LAAC. It consists of 0.025-inch and 0.035-inch magnet-tipped guidewires (FindrWIRZ) and 15-mm compliant occlusion balloon catheter (EndoCATH). It combines both endocardial and epicardial access. Computed tomographic (CT) screening is essential before the procedure to determine the suitability of LAA anatomy for the device. The contraindications for the device use include LAA diameter greater than 40 mm, prior cardiac surgery, pectus excavatum deformity, multilobed LAA, and posteriorly oriented LAA. However, the latest generation of LARIAT+ delivery devices has an increased snare width from 40 mm to 45 mm, enabling treatment for LAAs greater than 40 mm. The device received its CE-mark approval in 2015.

Sierra Litigation System

The Sierra Ligation System (Aegis Medical Innovations, Vancouver, Canada) is an epicardial device having single access through subxiphoid through electrographic navigation.[36] No transseptal puncture is required, unlike the Lariat system. The device consists of a LAAC grasper and a ligator, the former having articulating jaws with mounted electrodes to identify electric activity from the LAA.

Personnel

The procedure is performed in a standard cardiac catheterization laboratory. An anesthesiologist, an imager, and the implanter are the main participants in the procedure. The patient is given general sedation to protect the airway during extended TEE imaging and prevent potential sudden movements by the patient, which may cause cardiac perforation while manipulating LAA. The C-arm fluoroscopy is positioned just to the left of the operator, allowing more freedom of movement for the echocardiographer, who sits at the head of the table. Echocardiographers and anesthesiologists are provided with appropriate radiation shielding. The operator can see TEE images in real-time with the ultrasound machine slaved to the boom.[29][37]

Preparation

A pre-procedural transesophageal echocardiogram (TEE) is used for evaluating suitable candidates and determining the LAA morphology and dimensions. Important elements to be determined are the shape and size of the ostium, the width of the landing zone, the LAA length, and the number, shape, and location of the lobes, if possible.[38] The landing zone is defined as the area within the LAA where the device will be positioned. Additional pre-procedural imaging with computed tomography (CT) or magnetic resonance imaging (MRI) should be considered in the presence of large LAA neck width (greater than 26 mm) or a complicated LAA anatomy. Given size limitations associated with each device, good pre-procedural planning avoids unnecessary termination of the LAA closure procedure after the patient has undergone induction with general anesthesia.[39][37][40][41] 

Also, the LAA size depends on LA pressure and the presence of sinus rhythm or AF. Hence, certain operators prefer measuring LAA dimensions during the procedure after saline infusion establishes an LA pressure of greater than 10 mm Hg. To rule out post-procedural pericardial effusion and device embolization, many institutions and centers perform a chest radiograph and a TTE. To verify the complete sealing of the LAA and the absence of a thrombus, a TEE is recommended after 45 days. Post LAA closure, various antithrombotic/anticoagulation regimens have been used, including warfarin, dual antiplatelet therapy (DAPT), direct oral anticoagulants (DOAC), aspirin monotherapy, and no therapy. Therapy with either DAPT or DOAC does not significantly increase the incidence of device thrombus, stroke, or bleeding compared with standard warfarin therapy.[42] 

Technique or Treatment

Implantation of the device 

Percutaneous LAA closure is generally performed with general anesthesia guided by TEE and fluoroscopy. Prior to the procedure, antibiotic prophylaxis is usually given. Vascular access is achieved via a femoral vein, and transseptal puncture using a standard transseptal needle and sheath is done for the delivery system. The puncture site in the fossa is preferably inferior and posterior to have a good alignment with the LAA axis. Furthermore, there are different curves of sheaths available for accessing LAA. A bolus of unfractionated heparin should be provided after the puncture to obtain an activated clotting time (ACT) of more than 250 s. The size of the device is chosen based on LAA dimensions (ostium, neck width, depth) with the help of TEE and contrast angiography (right anterior oblique 30°/ cranial 30°). The desired implantation angulation of the C-arm may vary greatly from patient to patient. TEE and fluoroscopy ensure the location of the device in the LAA. The sheath over the device reduces the risk of perforation, and the device is subsequently deployed. A ‘tug test’ is performed once the device is in position to confirm device stability. Color Doppler imaging verifies complete sealing. The device is finally released from the delivery cable, and potential complications (e.g., pericardial effusion) are ruled out. The individual steps differ from one device to another and are described below.  

Watchman 

The device is selected by determining LAA size by the combination of TEE and fluoroscopic measurements. TEE excludes the presence of LAA thrombus before the procedure. However, cardiac CT can also be used for pre-procedural planning.[43] In the inferoposterior part of the interatrial septum, a 14-Fr double- or single- curved access sheath is inserted into the LA through a transseptal puncture. The 14-Fr sheath is placed deep inside the LAA and often telescoped over a diagnostic pigtail catheter to prevent laceration and perforation of LAA. The device is preloaded in a delivery system and is attached to a delivery cable. It is advanced to the tip of the access sheath, and the sheath is withdrawn. The device deployment follows this. TEE and fluoroscopy guide the device implantation, with TEE playing a critical role in determining the adequacy of the device employment. It is then released if the required TEE and fluoroscopic measurements are met by the counterclockwise rotation of the delivery cable.[44]  

Amplatzer Cardiac Plug and Amulet 

The landing zone is estimated between 12 mm to 15 mm from the LAA orifice both for ACP and Amulet.[45] The 12 Fr to 14 Fr double-curved delivery sheath is introduced into the LAA from the right femoral vein through a transseptal puncture. Post-procedural antiplatelet therapy is recommended for ACP for 1-6 months without OACs, and following it, aspirin is given indefinitely.[45] It is best to avoid appendages with more than 40 mm diameter, lobes behind the pulmonary artery, and posterior orientation of LAA.  

Lariat 

Cardiac CT determines the anatomic eligibility for the procedure. A second-generation device is larger and allows appendages up to 45 mm in diameter. The procedure requires an intact pericardium; hence it cannot be performed in patients with prior cardiac surgery. It is also discouraged in patients with prior pericarditis as the pericardial adhesions may prevent manipulation in the pericardial space and impede procedural success. Using an epidural or micropuncture needle under fluoroscopic guidance, a 14-Fr, slightly curved, and the soft-tipped sheath is advanced into the anterior pericardial space. Using the standard technique, after pericardial access is achieved, a transseptal sheath of 8.5-Fr is inserted into the LA from the right femoral vein; an inferoposterior location is usually favored in the interatrial septum. Through the transseptal sheath, a magnet-tipped guidewire is progressed into the anterior portion of the LAA, and a complementary magnet-tipped guidewire is progressed through the pericardial sheath and attached to the magnet-tipped wire in the LAA, forming a rail upon which the snare is delivered.

TEE and fluoroscopy guide the snare advancement into the pericardial sheath and over LAA, and LAA ostium is closed later. The snare is composed of a preloaded and tied surgical suture. Using a suture-tensioning device, the operator releases the suture with a proximal actuator and then tightens the prettied knot, after which the snare is withdrawn, and the suture is cut using a suture cutter. The pericardial sheath is substituted for a drain and remains in place for 4 to 6 hours, at least. The procedure can be summarized into four steps: Pericardial and transseptal puncture, endocardial magnet-tipped wire advancement into LAA apex, endocardial and epicardial magnet-tipped guidewires connection over which the Lariat snare is advanced, and the snare positioning and LAA ostium capture with the help of TEE-guided placement of inflated EndoCATH balloon catheter, and subsequently LAA suture ligation.[46]  

WaveCrest 

The WaveCrest is positioned to occlude the LAA, and its position is tested to confirm adequate seal and lack of a peri-device leak. The anchors are subsequently deployed to the distal edge of the device in a separate step. The "tug test" is subsequently performed to confirm the stability of the device before the device is released. Advantages include more flexibility while closing the LAA regardless of its depth and less traumatic deployment.[47]

Complications

The most common complications are pericardial effusion or cardiac tamponade requiring intervention, device-related thrombus, device embolization, persistent atrial septal defect (ASD), cardiac perforation, and procedure-related stroke. The incidence of pericardial effusion/tamponade is about 1.2% to 5% and, in the majority of cases, is managed with pericardiocentesis.[25][48][24][49][50][51][52][53][29][54] The incidence of device embolization has been reported as high as 3.5%, requiring transcatheter removal or surgery. However, the incidence is much lower amongst experienced operators (much less than 1%).[25][48][24][49][51][52][53][29][54] 

Device-related thrombus can be encountered in up to 14% of patients and is managed with a longer course of anticoagulation with a success rate approaching 95% (median treatment duration of 45 days).[55][48][51][52][53][29] The majority of these patients are asymptomatic, with an overall incidence of neurological events attributed to device-associated thrombus being very low (0.28%).[55] Persistent ASD incidence accounts for about 11% at six months and 7% at 12 months post-procedure.[56] Usually, ASD is small and does not require treatment. Cardiac perforations occur in up to 0.4% of patients and require surgical intervention.[25] Up to 1.1% of patients develop procedure-related stroke requiring management.[25][48][24][50][51][52][53][54] Other less common complications are procedure-related death and vascular complications such as hematoma, bleeding, and AV fistula formation.   

Clinical Significance

Clinical outcomes have been tested in various studies. Clinical efficacy and safety of the Watchman device have been evaluated in two randomized controlled trials (RCTs) and various large registries, demonstrating consistent improvements in procedural successes and complication rates. The PROTECT-AF trial and PREVAIL compared the WATCHMAN device and anticoagulation in patients with AF. The PROTECT-AF trial showed a procedural failure rate of 9%, with procedural-related complications of 8.7%, mostly secondary to pericardial effusion (4.8%).[24] This resulted in a high rate of primary safety events in the Watchman group compared to the anticoagulation group with a rate ratio (RR) of 1.69, 95% confidence interval (CI) of 1.01-3.19. The mean follow-up was 18 months. The primary endpoint was a stroke, systemic embolism, or cardiovascular death, which occurred in 3% in the study group versus 4.9% in control (RR-0.62, 95% credible interval, 0.35-1.25). Hemorrhagic strokes were less in the WATCHMAN group, as well as a reduction in the composite endpoint of death or disabling stroke in the device arm (RR-0.41, 95% CI, 0.22-0.82). There was about 60%, 34%, and 85% reduction in cardiovascular death (RR-0.4, 95% CI, 0.21-0.75), all-cause mortality (RR-0.66, 95% CI, 0.45-0.98) and hemorrhagic stroke (RR-0.15, 95% CI, 0.03-0.49).

Subsequently, the PREVAIL trial conducted on 407 patients showed a significant drop in procedure-related complications, even though it contained higher-risk patients, and about 40% of operators were novices.[25] It compared OAC-eligible patients with WATCHMAN device versus warfarin. The results showed a seven-day safety event rate of 2.2% in the Watchman group and a substantial reduction in serious procedural-related complications rate (4.2%, p = 0.004). Mean follow-up was for 12 months, and the non-inferiority criterion was not achieved for the first coprimary efficacy endpoint (stroke, systemic embolism, and cardiovascular death), but was achieved for the second coprimary endpoint (a stroke or systemic embolism greater than seven days; 0.02353 versus 0.02000 (risk difference, 0.0053; 95% credible interval, -0.0190 to 0.0273). Two large registries, the European EWOLUTION registry and the post-FDA approval US experience, included 1021 and 3822 patients, respectively. They demonstrated technical success rates of greater than 95% and significantly low rates of serious adverse effects within 7 days post-procedure.[54][52] Complication rates of serious pericardial effusion were 0.3% and 1.0%, respectively, with less than 0.5% of procedure-related stroke, embolization of device, and procedure-related death rates.  

ACP was first studied in 2011, demonstrating procedural success rates of 96% and serious adverse effects of 7%, of which 3.5% were pericardial effusion.[45] Another report from the largest multicenter registry studying 1047 patients showed a high procedural success rate of 97.3% with a serious adverse effects rate of 5%, of which cardiac tamponade and major bleeding accounted for 1.2%, ischemic stroke for 0.9%, and device embolization and procedural-related death of 0.8%.[57] Another prospective study of the Amulet device studied 1088 patients. It demonstrated technical success rates of 99% and procedural or in-hospital complication rates of 3.2%, of which pericardial tamponade accounted for 1.2%, procedure-related death of 0.2%, an ischemic stroke of 0.2%, and device embolization of 0.1%.[58] Another large observational registry with 500 patients showed device repositioning and left ventricular dysfunction associated with a higher number of serious adverse effects.[59] 

In patients with LAA anatomies unsuitable for endocardial devices, Lariat is a great option, although the procedure's safety is still a concern. There are many studies with the use of a Lariat device.[60][61][62][63][64][65] Results of a retrospective safety registry with 154 patients showed a high technical success rate of 94%.[60] However, major complications occurred in about 10%, significant pericardial effusion rate-10%, major bleeding rate-9%, and emergency surgery requirement-2% of the patients. High procedural complication rates, including death, have led to Lariat being confined to off-label use for LAAC based on FDA safety advisory recommendations. In another large series of the Lariat device in 712 patients showed a success rate of greater than 95% and complication rates of 5.3%, which was lower when compared to the prior studies. Complication rates were as follows: procedure-related death-0.1%, cardiac perforations requiring surgery-1.4%, and cardiac perforations not requiring surgery-2%. This decrease in cardiac perforation risk was due to the use of a micropuncture needle during pericardial access.

Enhancing Healthcare Team Outcomes

For AF patients, percutaneous LAAC has emerged as a valid alternative to OAC. Although randomized control trials have mostly been restricted to patients who can tolerate short-term OAC, registries have shown some benefit in patients at high risk of stroke and intolerance and/or contraindications to OAC, with consistent safety improvements as long-term efficacy of the procedure. Current recommendations from professional societies support the use of percutaneous LAAC in AF patients with high stroke risk and contraindication for long-term anticoagulation.[Level 1] Guidelines recommend 6 weeks of oral anticoagulation plus aspirin followed by aspirin plus clopidogrel up to 6 months post-procedure. Aspirin will subsequently be continued for the lifetime of the patient.[20][66] 

However, the post-procedural antithrombotic regimens are sometimes modified in select patients(especially in patients with recurrent bleeding after LAAC) to balance the risk of bleeding and device-associated thrombus. Patients undergo a transesophageal echocardiogram at 6 weeks to confirm adequate sealing of LAA and rule out device-related thrombus, as this might change the duration of oral anticoagulation. An interprofessional team including an interventional cardiologist, non-invasive cardiologist specialized in multimodality imaging, cardiac anesthesiologist, nurses, and catheterization laboratory technologists work together to provide an integrated and holistic approach to pre-procedural planning, intra-procedural support, and post-procedural care to ensure an optimal outcome for the patient. Communication, shared decision-making, and collaboration between the interprofessional team members are key elements for excellent outcomes. Interprofessional care also helps identify early and late complications of percutaneous LAAC and, therefore, improves patients' prognosis and improves patient outcomes while reducing adverse events. [Level 5]


Details

Editor:

Reza Arsanjani

Updated:

6/12/2023 7:57:40 PM

References


[1]

Go AS, Hylek EM, Phillips KA, Chang Y, Henault LE, Selby JV, Singer DE. Prevalence of diagnosed atrial fibrillation in adults: national implications for rhythm management and stroke prevention: the AnTicoagulation and Risk Factors in Atrial Fibrillation (ATRIA) Study. JAMA. 2001 May 9:285(18):2370-5     [PubMed PMID: 11343485]


[2]

Chugh SS, Havmoeller R, Narayanan K, Singh D, Rienstra M, Benjamin EJ, Gillum RF, Kim YH, McAnulty JH Jr, Zheng ZJ, Forouzanfar MH, Naghavi M, Mensah GA, Ezzati M, Murray CJ. Worldwide epidemiology of atrial fibrillation: a Global Burden of Disease 2010 Study. Circulation. 2014 Feb 25:129(8):837-47. doi: 10.1161/CIRCULATIONAHA.113.005119. Epub 2013 Dec 17     [PubMed PMID: 24345399]


[3]

Benjamin EJ, Blaha MJ, Chiuve SE, Cushman M, Das SR, Deo R, de Ferranti SD, Floyd J, Fornage M, Gillespie C, Isasi CR, Jiménez MC, Jordan LC, Judd SE, Lackland D, Lichtman JH, Lisabeth L, Liu S, Longenecker CT, Mackey RH, Matsushita K, Mozaffarian D, Mussolino ME, Nasir K, Neumar RW, Palaniappan L, Pandey DK, Thiagarajan RR, Reeves MJ, Ritchey M, Rodriguez CJ, Roth GA, Rosamond WD, Sasson C, Towfighi A, Tsao CW, Turner MB, Virani SS, Voeks JH, Willey JZ, Wilkins JT, Wu JH, Alger HM, Wong SS, Muntner P, American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Heart Disease and Stroke Statistics-2017 Update: A Report From the American Heart Association. Circulation. 2017 Mar 7:135(10):e146-e603. doi: 10.1161/CIR.0000000000000485. Epub 2017 Jan 25     [PubMed PMID: 28122885]


[4]

Wolf PA, Abbott RD, Kannel WB. Atrial fibrillation as an independent risk factor for stroke: the Framingham Study. Stroke. 1991 Aug:22(8):983-8     [PubMed PMID: 1866765]


[5]

Frost L, Engholm G, Johnsen S, Møller H, Henneberg EW, Husted S. Incident thromboembolism in the aorta and the renal, mesenteric, pelvic, and extremity arteries after discharge from the hospital with a diagnosis of atrial fibrillation. Archives of internal medicine. 2001 Jan 22:161(2):272-6     [PubMed PMID: 11176743]


[6]

Wang TJ, Massaro JM, Levy D, Vasan RS, Wolf PA, D'Agostino RB, Larson MG, Kannel WB, Benjamin EJ. A risk score for predicting stroke or death in individuals with new-onset atrial fibrillation in the community: the Framingham Heart Study. JAMA. 2003 Aug 27:290(8):1049-56     [PubMed PMID: 12941677]


[7]

Hart RG, Benavente O, McBride R, Pearce LA. Antithrombotic therapy to prevent stroke in patients with atrial fibrillation: a meta-analysis. Annals of internal medicine. 1999 Oct 5:131(7):492-501     [PubMed PMID: 10507957]

Level 1 (high-level) evidence

[8]

. Risk factors for stroke and efficacy of antithrombotic therapy in atrial fibrillation. Analysis of pooled data from five randomized controlled trials. Archives of internal medicine. 1994 Jul 11:154(13):1449-57     [PubMed PMID: 8018000]

Level 1 (high-level) evidence

[9]

Al-Saady NM, Obel OA, Camm AJ. Left atrial appendage: structure, function, and role in thromboembolism. Heart (British Cardiac Society). 1999 Nov:82(5):547-54     [PubMed PMID: 10525506]


[10]

Wunderlich NC, Beigel R, Swaans MJ, Ho SY, Siegel RJ. Percutaneous interventions for left atrial appendage exclusion: options, assessment, and imaging using 2D and 3D echocardiography. JACC. Cardiovascular imaging. 2015 Apr:8(4):472-488. doi: 10.1016/j.jcmg.2015.02.002. Epub     [PubMed PMID: 25882576]


[11]

Shirani J, Alaeddini J. Structural remodeling of the left atrial appendage in patients with chronic non-valvular atrial fibrillation: Implications for thrombus formation, systemic embolism, and assessment by transesophageal echocardiography. Cardiovascular pathology : the official journal of the Society for Cardiovascular Pathology. 2000 Mar-Apr:9(2):95-101     [PubMed PMID: 10867359]


[12]

Manning WJ, Silverman DI, Keighley CS, Oettgen P, Douglas PS. Transesophageal echocardiographically facilitated early cardioversion from atrial fibrillation using short-term anticoagulation: final results of a prospective 4.5-year study. Journal of the American College of Cardiology. 1995 May:25(6):1354-61     [PubMed PMID: 7722133]


[13]

Leung DY, Black IW, Cranney GB, Hopkins AP, Walsh WF. Prognostic implications of left atrial spontaneous echo contrast in nonvalvular atrial fibrillation. Journal of the American College of Cardiology. 1994 Sep:24(3):755-62     [PubMed PMID: 8077549]


[14]

January CT, Wann LS, Alpert JS, Calkins H, Cigarroa JE, Cleveland JC Jr, Conti JB, Ellinor PT, Ezekowitz MD, Field ME, Murray KT, Sacco RL, Stevenson WG, Tchou PJ, Tracy CM, Yancy CW, American College of Cardiology/American Heart Association Task Force on Practice Guidelines. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. Journal of the American College of Cardiology. 2014 Dec 2:64(21):e1-76. doi: 10.1016/j.jacc.2014.03.022. Epub 2014 Mar 28     [PubMed PMID: 24685669]

Level 1 (high-level) evidence

[15]

Klein AL, Grimm RA, Murray RD, Apperson-Hansen C, Asinger RW, Black IW, Davidoff R, Erbel R, Halperin JL, Orsinelli DA, Porter TR, Stoddard MF, Assessment of Cardioversion Using Transesophageal Echocardiography Investigators. Use of transesophageal echocardiography to guide cardioversion in patients with atrial fibrillation. The New England journal of medicine. 2001 May 10:344(19):1411-20     [PubMed PMID: 11346805]


[16]

Blackshear JL, Odell JA. Appendage obliteration to reduce stroke in cardiac surgical patients with atrial fibrillation. The Annals of thoracic surgery. 1996 Feb:61(2):755-9     [PubMed PMID: 8572814]


[17]

. Predictors of thromboembolism in atrial fibrillation: II. Echocardiographic features of patients at risk. The Stroke Prevention in Atrial Fibrillation Investigators. Annals of internal medicine. 1992 Jan 1:116(1):6-12     [PubMed PMID: 1727097]


[18]

January CT, Wann LS, Calkins H, Chen LY, Cigarroa JE, Cleveland JC Jr, Ellinor PT, Ezekowitz MD, Field ME, Furie KL, Heidenreich PA, Murray KT, Shea JB, Tracy CM, Yancy CW. 2019 AHA/ACC/HRS Focused Update of the 2014 AHA/ACC/HRS Guideline for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society in Collaboration With the Society of Thoracic Surgeons. Circulation. 2019 Jul 9:140(2):e125-e151. doi: 10.1161/CIR.0000000000000665. Epub 2019 Jan 28     [PubMed PMID: 30686041]

Level 1 (high-level) evidence

[19]

Kirchhof P, Benussi S, Kotecha D, Ahlsson A, Atar D, Casadei B, Castella M, Diener HC, Heidbuchel H, Hendriks J, Hindricks G, Manolis AS, Oldgren J, Popescu BA, Schotten U, Van Putte B, Vardas P, ESC Scientific Document Group. 2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS. European heart journal. 2016 Oct 7:37(38):2893-2962. doi: 10.1093/eurheartj/ehw210. Epub 2016 Aug 27     [PubMed PMID: 27567408]


[20]

Writing Group Members, January CT, Wann LS, Calkins H, Chen LY, Cigarroa JE, Cleveland JC Jr, Ellinor PT, Ezekowitz MD, Field ME, Furie KL, Heidenreich PA, Murray KT, Shea JB, Tracy CM, Yancy CW. 2019 AHA/ACC/HRS focused update of the 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society. Heart rhythm. 2019 Aug:16(8):e66-e93. doi: 10.1016/j.hrthm.2019.01.024. Epub 2019 Jan 28     [PubMed PMID: 30703530]

Level 1 (high-level) evidence

[21]

Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, Hindricks G, Kirchhof P, ESC Committee for Practice Guidelines-CPG, Document Reviewers. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation--developed with the special contribution of the European Heart Rhythm Association. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2012 Oct:14(10):1385-413     [PubMed PMID: 22923145]


[22]

Sievert H, Lesh MD, Trepels T, Omran H, Bartorelli A, Della Bella P, Nakai T, Reisman M, DiMario C, Block P, Kramer P, Fleschenberg D, Krumsdorf U, Scherer D. Percutaneous left atrial appendage transcatheter occlusion to prevent stroke in high-risk patients with atrial fibrillation: early clinical experience. Circulation. 2002 Apr 23:105(16):1887-9     [PubMed PMID: 11997272]


[23]

Bayard YL, Omran H, Neuzil P, Thuesen L, Pichler M, Rowland E, Ramondo A, Ruzyllo W, Budts W, Montalescot G, Brugada P, Serruys PW, Vahanian A, Piéchaud JF, Bartorelli A, Marco J, Probst P, Kuck KH, Ostermayer SH, Büscheck F, Fischer E, Leetz M, Sievert H. PLAATO (Percutaneous Left Atrial Appendage Transcatheter Occlusion) for prevention of cardioembolic stroke in non-anticoagulation eligible atrial fibrillation patients: results from the European PLAATO study. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2010 Jun:6(2):220-6     [PubMed PMID: 20562072]


[24]

Holmes DR, Reddy VY, Turi ZG, Doshi SK, Sievert H, Buchbinder M, Mullin CM, Sick P, PROTECT AF Investigators. Percutaneous closure of the left atrial appendage versus warfarin therapy for prevention of stroke in patients with atrial fibrillation: a randomised non-inferiority trial. Lancet (London, England). 2009 Aug 15:374(9689):534-42. doi: 10.1016/S0140-6736(09)61343-X. Epub     [PubMed PMID: 19683639]

Level 1 (high-level) evidence

[25]

Holmes DR Jr, Kar S, Price MJ, Whisenant B, Sievert H, Doshi SK, Huber K, Reddy VY. Prospective randomized evaluation of the Watchman Left Atrial Appendage Closure device in patients with atrial fibrillation versus long-term warfarin therapy: the PREVAIL trial. Journal of the American College of Cardiology. 2014 Jul 8:64(1):1-12. doi: 10.1016/j.jacc.2014.04.029. Epub     [PubMed PMID: 24998121]

Level 1 (high-level) evidence

[26]

Reddy VY, Doshi SK, Sievert H, Buchbinder M, Neuzil P, Huber K, Halperin JL, Holmes D, PROTECT AF Investigators. Percutaneous left atrial appendage closure for stroke prophylaxis in patients with atrial fibrillation: 2.3-Year Follow-up of the PROTECT AF (Watchman Left Atrial Appendage System for Embolic Protection in Patients with Atrial Fibrillation) Trial. Circulation. 2013 Feb 12:127(6):720-9. doi: 10.1161/CIRCULATIONAHA.112.114389. Epub 2013 Jan 16     [PubMed PMID: 23325525]


[27]

Seeger J, Birkemeyer R, Rottbauer W, Wöhrle J. First experience with the Watchman FLX occluder for percutaneous left atrial appendage closure. Cardiovascular revascularization medicine : including molecular interventions. 2017 Oct-Nov:18(7):512-516. doi: 10.1016/j.carrev.2017.04.018. Epub 2017 Apr 29     [PubMed PMID: 28495117]


[28]

Swaans MJ, Wintgens LI, Alipour A, Rensing BJ, Boersma LV. Percutaneous left atrial appendage closure devices: safety, efficacy, and clinical utility. Medical devices (Auckland, N.Z.). 2016:9():309-16. doi: 10.2147/MDER.S65492. Epub 2016 Sep 2     [PubMed PMID: 27621674]


[29]

Tzikas A, Gafoor S, Meerkin D, Freixa X, Cruz-Gonzalez I, Lewalter T, Saw J, Berti S, Nielsen-Kudsk JE, Ibrahim R, Lakkireddy D, Paul V, Arzamendi D, Nietlispach F, Worthley SG, Hildick-Smith D, Thambo JB, Tondo C, Aminian A, Kalarus Z, Schmidt B, Sondergaard L, Kefer J, Meier B, Park JW, Sievert H, Omran H. Left atrial appendage occlusion with the AMPLATZER Amulet device: an expert consensus step-by-step approach. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2016 Apr 20:11(13):1512-21. doi: 10.4244/EIJV11I13A292. Epub     [PubMed PMID: 27107315]

Level 3 (low-level) evidence

[30]

De Backer O, Arnous S, Ihlemann N, Vejlstrup N, Jørgensen E, Pehrson S, Krieger TD, Meier P, Søndergaard L, Franzen OW. Percutaneous left atrial appendage occlusion for stroke prevention in atrial fibrillation: an update. Open heart. 2014:1(1):e000020. doi: 10.1136/openhrt-2013-000020. Epub 2014 Jun 6     [PubMed PMID: 25332785]


[31]

Bellmann B, Schnupp S, Kühnlein P, Javernik C, Kleinecke C, Rillig A, Landmesser U, Brachmann J, Park JW. Left Atrial Appendage Closure With the New Occlutech® Device: First in Man Experience and Neurological Outcome. Journal of cardiovascular electrophysiology. 2017 Mar:28(3):315-320. doi: 10.1111/jce.13141. Epub 2016 Dec 29     [PubMed PMID: 27925332]


[32]

Cruz-González I, Freixa X, Fernández-Díaz JA, Moreno-Samos JC, Martín-Yuste V, Goicolea J. Left Atrial Appendage Occlusion With the LAmbre Device: Initial Experience. Revista espanola de cardiologia (English ed.). 2018 Sep:71(9):755-756. doi: 10.1016/j.rec.2017.04.015. Epub 2017 May 15     [PubMed PMID: 28522304]


[33]

Regueiro A, Bernier M, O'Hara G, O'Connor K, Paradis JM, Beaudoin J, Rodriguez-Gabella T, Champagne J, Rodés-Cabau J. Left atrial appendage closure: Initial experience with the ultraseal device. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2017 Nov 1:90(5):817-823. doi: 10.1002/ccd.26870. Epub 2016 Nov 17     [PubMed PMID: 27860249]


[34]

Asmarats L, Rodés-Cabau J. Percutaneous Left Atrial Appendage Closure: Current Devices and Clinical Outcomes. Circulation. Cardiovascular interventions. 2017 Nov:10(11):. pii: e005359. doi: 10.1161/CIRCINTERVENTIONS.117.005359. Epub     [PubMed PMID: 29146668]

Level 2 (mid-level) evidence

[35]

Lee RJ, Bartus K, Yakubov SJ. Catheter-based left atrial appendage (LAA) ligation for the prevention of embolic events arising from the LAA: initial experience in a canine model. Circulation. Cardiovascular interventions. 2010 Jun 1:3(3):224-9. doi: 10.1161/CIRCINTERVENTIONS.109.914978. Epub 2010 May 18     [PubMed PMID: 20484100]


[36]

Bruce CJ, Stanton CM, Asirvatham SJ, Danielsen AJ, Johnson SB, Packer DL, Friedman PA. Percutaneous epicardial left atrial appendage closure: intermediate-term results. Journal of cardiovascular electrophysiology. 2011 Jan:22(1):64-70. doi: 10.1111/j.1540-8167.2010.01855.x. Epub     [PubMed PMID: 20662983]


[37]

Budge LP, Shaffer KM, Moorman JR, Lake DE, Ferguson JD, Mangrum JM. Analysis of in vivo left atrial appendage morphology in patients with atrial fibrillation: a direct comparison of transesophageal echocardiography, planar cardiac CT, and segmented three-dimensional cardiac CT. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2008 Nov:23(2):87-93. doi: 10.1007/s10840-008-9281-7. Epub 2008 Aug 7     [PubMed PMID: 18686024]


[38]

Chue CD, de Giovanni J, Steeds RP. The role of echocardiography in percutaneous left atrial appendage occlusion. European journal of echocardiography : the journal of the Working Group on Echocardiography of the European Society of Cardiology. 2011 Oct:12(10):i3-10. doi: 10.1093/ejechocard/jer090. Epub     [PubMed PMID: 21998466]


[39]

Wang Y, Di Biase L, Horton RP, Nguyen T, Morhanty P, Natale A. Left atrial appendage studied by computed tomography to help planning for appendage closure device placement. Journal of cardiovascular electrophysiology. 2010 Sep:21(9):973-82. doi: 10.1111/j.1540-8167.2010.01814.x. Epub     [PubMed PMID: 20550614]


[40]

Krishnaswamy A, Patel NS, Ozkan A, Agarwal S, Griffin BP, Saliba W, Tuzcu EM, Schoenhagen P, Kapadia SR. Planning left atrial appendage occlusion using cardiac multidetector computed tomography. International journal of cardiology. 2012 Jul 12:158(2):313-7. doi: 10.1016/j.ijcard.2012.04.147. Epub 2012 May 21     [PubMed PMID: 22622059]


[41]

Heist EK, Refaat M, Danik SB, Holmvang G, Ruskin JN, Mansour M. Analysis of the left atrial appendage by magnetic resonance angiography in patients with atrial fibrillation. Heart rhythm. 2006 Nov:3(11):1313-8     [PubMed PMID: 17074637]


[42]

Bergmann MW, Betts TR, Sievert H, Schmidt B, Pokushalov E, Kische S, Schmitz T, Meincke F, Stein KM, Boersma LVA, Ince H. Safety and efficacy of early anticoagulation drug regimens after WATCHMAN left atrial appendage closure: three-month data from the EWOLUTION prospective, multicentre, monitored international WATCHMAN LAA closure registry. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2017 Sep 20:13(7):877-884. doi: 10.4244/EIJ-D-17-00042. Epub     [PubMed PMID: 28606886]


[43]

Wang DD, Eng M, Kupsky D, Myers E, Forbes M, Rahman M, Zaidan M, Parikh S, Wyman J, Pantelic M, Song T, Nadig J, Karabon P, Greenbaum A, O'Neill W. Application of 3-Dimensional Computed Tomographic Image Guidance to WATCHMAN Implantation and Impact on Early Operator Learning Curve: Single-Center Experience. JACC. Cardiovascular interventions. 2016 Nov 28:9(22):2329-2340. doi: 10.1016/j.jcin.2016.07.038. Epub     [PubMed PMID: 27884358]


[44]

Saw J, Lempereur M. Percutaneous left atrial appendage closure: procedural techniques and outcomes. JACC. Cardiovascular interventions. 2014 Nov:7(11):1205-20. doi: 10.1016/j.jcin.2014.05.026. Epub 2014 Nov 17     [PubMed PMID: 25459035]


[45]

Park JW, Bethencourt A, Sievert H, Santoro G, Meier B, Walsh K, Lopez-Minguez JR, Meerkin D, Valdés M, Ormerod O, Leithäuser B. Left atrial appendage closure with Amplatzer cardiac plug in atrial fibrillation: initial European experience. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2011 Apr 1:77(5):700-6. doi: 10.1002/ccd.22764. Epub 2011 Mar 8     [PubMed PMID: 20824765]


[46]

Janga P, Pamulapati H, Kanmanthareddy A, Vallakati A, Gunda S, Bommana S, Jaeger M, Yeruva MR, Lakkireddy D. State Of The Art In Left Atrial Appendage Ligation - The Lariat. Journal of atrial fibrillation. 2014 Jun-Jul:7(1):1073. doi: 10.4022/jafib.1073. Epub 2014 Jun 30     [PubMed PMID: 27957084]


[47]

Vainrib AF, Bamira D, Benenstein RJ, Aizer A, Chinitz LA, Saric M. Echocardiographic Guidance of the Novel WaveCrest Left Atrial Appendage Occlusion Device. CASE (Philadelphia, Pa.). 2018 Dec:2(6):297-300. doi: 10.1016/j.case.2018.07.005. Epub 2018 Sep 27     [PubMed PMID: 30582095]

Level 3 (low-level) evidence

[48]

Sick PB, Schuler G, Hauptmann KE, Grube E, Yakubov S, Turi ZG, Mishkel G, Almany S, Holmes DR. Initial worldwide experience with the WATCHMAN left atrial appendage system for stroke prevention in atrial fibrillation. Journal of the American College of Cardiology. 2007 Apr 3:49(13):1490-5     [PubMed PMID: 17397680]


[49]

Reddy VY, Sievert H, Halperin J, Doshi SK, Buchbinder M, Neuzil P, Huber K, Whisenant B, Kar S, Swarup V, Gordon N, Holmes D, PROTECT AF Steering Committee and Investigators. Percutaneous left atrial appendage closure vs warfarin for atrial fibrillation: a randomized clinical trial. JAMA. 2014 Nov 19:312(19):1988-98. doi: 10.1001/jama.2014.15192. Epub     [PubMed PMID: 25399274]

Level 1 (high-level) evidence

[50]

Reddy VY, Holmes D, Doshi SK, Neuzil P, Kar S. Safety of percutaneous left atrial appendage closure: results from the Watchman Left Atrial Appendage System for Embolic Protection in Patients with AF (PROTECT AF) clinical trial and the Continued Access Registry. Circulation. 2011 Feb 1:123(4):417-24. doi: 10.1161/CIRCULATIONAHA.110.976449. Epub 2011 Jan 17     [PubMed PMID: 21242484]


[51]

Boersma LV, Ince H, Kische S, Pokushalov E, Schmitz T, Schmidt B, Gori T, Meincke F, Protopopov AV, Betts T, Foley D, Sievert H, Mazzone P, De Potter T, Vireca E, Stein K, Bergmann MW, EWOLUTION Investigators. Efficacy and safety of left atrial appendage closure with WATCHMAN in patients with or without contraindication to oral anticoagulation: 1-Year follow-up outcome data of the EWOLUTION trial. Heart rhythm. 2017 Sep:14(9):1302-1308. doi: 10.1016/j.hrthm.2017.05.038. Epub 2017 May 31     [PubMed PMID: 28577840]


[52]

Boersma LV, Schmidt B, Betts TR, Sievert H, Tamburino C, Teiger E, Pokushalov E, Kische S, Schmitz T, Stein KM, Bergmann MW, EWOLUTION investigators. Implant success and safety of left atrial appendage closure with the WATCHMAN device: peri-procedural outcomes from the EWOLUTION registry. European heart journal. 2016 Aug:37(31):2465-74. doi: 10.1093/eurheartj/ehv730. Epub 2016 Jan 27     [PubMed PMID: 26822918]


[53]

Reddy VY, Möbius-Winkler S, Miller MA, Neuzil P, Schuler G, Wiebe J, Sick P, Sievert H. Left atrial appendage closure with the Watchman device in patients with a contraindication for oral anticoagulation: the ASAP study (ASA Plavix Feasibility Study With Watchman Left Atrial Appendage Closure Technology). Journal of the American College of Cardiology. 2013 Jun 25:61(25):2551-6. doi: 10.1016/j.jacc.2013.03.035. Epub 2013 Apr 10     [PubMed PMID: 23583249]

Level 2 (mid-level) evidence

[54]

Reddy VY, Gibson DN, Kar S, O'Neill W, Doshi SK, Horton RP, Buchbinder M, Gordon NT, Holmes DR. Post-Approval U.S. Experience With Left Atrial Appendage Closure for Stroke Prevention in Atrial Fibrillation. Journal of the American College of Cardiology. 2017 Jan 24:69(3):253-261. doi: 10.1016/j.jacc.2016.10.010. Epub 2016 Nov 2     [PubMed PMID: 27816552]


[55]

Lempereur M, Aminian A, Freixa X, Gafoor S, Kefer J, Tzikas A, Legrand V, Saw J. Device-associated thrombus formation after left atrial appendage occlusion: A systematic review of events reported with the Watchman, the Amplatzer Cardiac Plug and the Amulet. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2017 Nov 1:90(5):E111-E121. doi: 10.1002/ccd.26903. Epub 2017 Feb 1     [PubMed PMID: 28145040]

Level 1 (high-level) evidence

[56]

Singh SM, Douglas PS, Reddy VY. The incidence and long-term clinical outcome of iatrogenic atrial septal defects secondary to transseptal catheterization with a 12F transseptal sheath. Circulation. Arrhythmia and electrophysiology. 2011 Apr:4(2):166-71. doi: 10.1161/CIRCEP.110.959015. Epub 2011 Jan 19     [PubMed PMID: 21248245]

Level 2 (mid-level) evidence

[57]

Tzikas A, Shakir S, Gafoor S, Omran H, Berti S, Santoro G, Kefer J, Landmesser U, Nielsen-Kudsk JE, Cruz-Gonzalez I, Sievert H, Tichelbäcker T, Kanagaratnam P, Nietlispach F, Aminian A, Kasch F, Freixa X, Danna P, Rezzaghi M, Vermeersch P, Stock F, Stolcova M, Costa M, Ibrahim R, Schillinger W, Meier B, Park JW. Left atrial appendage occlusion for stroke prevention in atrial fibrillation: multicentre experience with the AMPLATZER Cardiac Plug. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2016 Feb:11(10):1170-9. doi: 10.4244/EIJY15M01_06. Epub     [PubMed PMID: 25604089]


[58]

Landmesser U, Schmidt B, Nielsen-Kudsk JE, Lam SCC, Park JW, Tarantini G, Cruz-Gonzalez I, Geist V, Della Bella P, Colombo A, Zeus T, Omran H, Piorkowski C, Lund J, Tondo C, Hildick-Smith D. Left atrial appendage occlusion with the AMPLATZER Amulet device: periprocedural and early clinical/echocardiographic data from a global prospective observational study. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2017 Sep 20:13(7):867-876. doi: 10.4244/EIJ-D-17-00493. Epub     [PubMed PMID: 28649053]

Level 2 (mid-level) evidence

[59]

Koskinas KC, Shakir S, Fankhauser M, Nietlispach F, Attinger-Toller A, Moschovitis A, Wenaweser P, Pilgrim T, Stortecky S, Praz F, Räber L, Windecker S, Meier B, Gloekler S. Predictors of Early (1-Week) Outcomes Following Left Atrial Appendage Closure With Amplatzer Devices. JACC. Cardiovascular interventions. 2016 Jul 11:9(13):1374-83. doi: 10.1016/j.jcin.2016.04.019. Epub     [PubMed PMID: 27388826]


[60]

Price MJ, Gibson DN, Yakubov SJ, Schultz JC, Di Biase L, Natale A, Burkhardt JD, Pershad A, Byrne TJ, Gidney B, Aragon JR, Goldstein J, Moulton K, Patel T, Knight B, Lin AC, Valderrábano M. Early safety and efficacy of percutaneous left atrial appendage suture ligation: results from the U.S. transcatheter LAA ligation consortium. Journal of the American College of Cardiology. 2014 Aug 12:64(6):565-72. doi: 10.1016/j.jacc.2014.03.057. Epub     [PubMed PMID: 25104525]


[61]

Lakkireddy D, Afzal MR, Lee RJ, Nagaraj H, Tschopp D, Gidney B, Ellis C, Altman E, Lee B, Kar S, Bhadwar N, Sanchez M, Gadiyaram V, Evonich R, Rasekh A, Cheng J, Cuoco F, Chandhok S, Gunda S, Reddy M, Atkins D, Bommana S, Cuculich P, Gibson D, Nath J, Ferrell R, Matthew E, Wilber D. Short and long-term outcomes of percutaneous left atrial appendage suture ligation: Results from a US multicenter evaluation. Heart rhythm. 2016 May:13(5):1030-1036. doi: 10.1016/j.hrthm.2016.01.022. Epub 2016 Feb 9     [PubMed PMID: 26872554]


[62]

Matsuo Y, Sandri M, Mangner N, Majunke N, Dähnert I, Schuler G, Kurabayashi M, Möbius-Winkler S. Interventional closure of the left atrial appendage for stroke prevention. Circulation journal : official journal of the Japanese Circulation Society. 2014:78(3):619-24     [PubMed PMID: 24419803]


[63]

Kim JS, Lee H, Suh Y, Pak HN, Hong GR, Shim CY, Yu CW, Lee HJ, Kang WC, Shin ES, Choi RK, Kar S, Park JW, Lim DS, Jang Y. Left Atrial Appendage Occlusion in Non-Valvular Atrial Fibrillation in a Korean Multi-Center Registry. Circulation journal : official journal of the Japanese Circulation Society. 2016 Apr 25:80(5):1123-30. doi: 10.1253/circj.CJ-15-1134. Epub 2016 Mar 17     [PubMed PMID: 26984716]


[64]

Figini F, Mazzone P, Regazzoli D, Porata G, Ruparelia N, Giannini F, Stella S, Ancona F, Agricola E, Sora N, Marzi A, Aurelio A, Trevisi N, Della Bella P, Colombo A, Montorfano M. Left atrial appendage closure: A single center experience and comparison of two contemporary devices. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2017 Mar 1:89(4):763-772. doi: 10.1002/ccd.26678. Epub 2016 Aug 27     [PubMed PMID: 27567013]


[65]

Betts TR, Leo M, Panikker S, Kanagaratnam P, Koa-Wing M, Davies DW, Hildick-Smith D, Wynne DG, Ormerod O, Segal OR, Chow AW, Todd D, Cabrera Gomez S, Kirkwood GJ, Fox D, Pepper C, Foran J, Wong T. Percutaneous left atrial appendage occlusion using different technologies in the United Kingdom: A multicenter registry. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2017 Feb 15:89(3):484-492. doi: 10.1002/ccd.26782. Epub 2016 Sep 21     [PubMed PMID: 27651124]


[66]

Kirchhof P, Benussi S, Kotecha D, Ahlsson A, Atar D, Casadei B, Castella M, Diener HC, Heidbuchel H, Hendriks J, Hindricks G, Manolis AS, Oldgren J, Popescu BA, Schotten U, Van Putte B, Vardas P, Agewall S, Camm J, Baron Esquivias G, Budts W, Carerj S, Casselman F, Coca A, De Caterina R, Deftereos S, Dobrev D, Ferro JM, Filippatos G, Fitzsimons D, Gorenek B, Guenoun M, Hohnloser SH, Kolh P, Lip GY, Manolis A, McMurray J, Ponikowski P, Rosenhek R, Ruschitzka F, Savelieva I, Sharma S, Suwalski P, Tamargo JL, Taylor CJ, Van Gelder IC, Voors AA, Windecker S, Zamorano JL, Zeppenfeld K. 2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2016 Nov:18(11):1609-1678     [PubMed PMID: 27567465]